Diagnostic kit (Colloidal Gold) for IgG/IgM antibody to dengue is an in vitro colloidal gold assay for the qualitative detection of IgG/IgM antibody to dengue in serum/plasma/whole blood samples. It is intended to aid in the rapid diagnosis of dengue virus infections. The result can be obtained in 15 minutes.

Intended Use

Dengue fever is an acute infectious disease caused by viruses, which has the symptoms of double phase fever, headache and pain in different parts of body. Our IgG/IgM dengue fever test kit is used for in vitro qualitative detection of IgG/IgM antibody to dengue in human whole blood, serum or plasma sample, which's applicable to auxiliary diagnosis of dengue virus infection. This kit is for healthcare professionals.


Benefits

  • Rapid: The time to result is in 15 minutes, which helps you accelerate diagnosis process.

  • Cost-effective: There is no additional machine needed to run the test, you don't need to worried about machine cost.

  • Screening suitable: The easy test procedure ensure you the possiblity take our dengue fever test kit as a screening tool.


Product Specifications

Method

Colloidal Gold

Sample Type

Whole blood, serum, plasma

Time to Result

15-20 mins

Storge

2~30/36~86℉

Shelf Life

24 months from date of manufacture

Kit Size

1/5/20/25 tests

Refer to Package Insert for additional product information.


Product Performance

WIZ results of dengue IgG antibody

Test results of reference reagents

Positive coincidence rate:

91.33% (95%C.I. 86.55%~94.51%)


Negative coincidence rate:

97.01% (95%C.I. 93.18%~98.71%)


Total coincidence rate:

93.94% (95%C.I. 90.99%~95.96%)

POS

NEG

Total

POS

179

5

184

NEG

17

162

179

Total

196

167

363


WIZ results of dengue IgM antibody

Test results of reference reagents

Positive coincidence rate:

91.06% (95%C.I. 85.97%~94.42%)


Negative coincidence rate:

97.83% (95%C.I. 94.54%~99.15%)


Total coincidence rate:

94.49% (95%C.I. 91.64%~96.41%)

POS

NEG

Total

POS

163

4

167

NEG

16

180

196

Total

179

184

363